EasyManua.ls Logo

R&S SMA100B - Page 558

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
558User Manual 1178.3834.02 ─ 09
:SENSe<ch>[:POWer]:LOGGing:STATe <State>
Activates the recording of the power values, measured by a connected R&S NRP
power sensor.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENS:LOGG:STAT ON
Activates recording of the power measurement of the first sen-
sor.
Manual operation: See "Enable Logging" on page 244
:SENSe<ch>[:POWer]:OFFSet <Offset>
Sets a level offset which is added to the measured level value after activation with
command :SENSe<ch>[:POWer]:OFFSet:STATe. The level offset allows, e.g. to
consider an attenuator in the signal path.
Parameters:
<Offset> float
Range: -100.0 to 100.0
*RST: 0
Default unit: dB
Example:
:SENS1:POW:OFFS 10.0
Sets a level offset of 10 dB
Manual operation: See "Level Offset State,Level Offset" on page 242
:SENSe<ch>[:POWer]:OFFSet:STATe <State>
Activates the addition of the level offset to the measured value. The level offset value is
set with command :SENSe<ch>[:POWer]:OFFSet.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENS1:POW:OFFS 0.4dB
Sets a level offset of 0.4 dB
:SENS1:POW:OFFS:STAT ON
A level offset of 0.4 dB is added to the measured value.
Manual operation: See "Level Offset State,Level Offset" on page 242
:SENSe<ch>[:POWer]:SNUMber?
Queries the serial number of the sensor.
SENSe, READ, INITiate and SLISt subsystems

Table of Contents

Other manuals for R&S SMA100B

Related product manuals