Remote control commands
R&S
®
SMA100B
559User Manual 1178.3834.02 ─ 09
Return values:
<SNumber> string
Example:
:SENS1:SNUM?
Queries the serial number.
Usage: Query only
Manual operation: See "Sensor type and serial number" on page 241
:SENSe<ch>[:POWer]:SOURce <Source>
Determines the signal to be measured.
Note: When measuring the RF signal, the sensor considers the corresponding correc-
tion factor at that frequency, and uses the level setting of the instrument as reference
level.
Parameters:
<Source> A | USER | RF
*RST: A
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 548.
Manual operation: See "Use Frequency Of" on page 242
:SENSe<ch>[:POWer]:STATus[:DEVice]?
Queries if a sensor is connected to the instrument.
Return values:
<Status> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENS1:STAT?
Response: 1
A sensor is connected.
Usage: Query only
Manual operation: See "State" on page 241
:SENSe<ch>[:POWer]:SVERsion?
Queries the software version of the connected R&S NRP power sensor.
Return values:
<SVersion> string
Example:
:SENS1:POW:SVER?
Queries the software version of the power sensor.
Usage: Query only
SENSe, READ, INITiate and SLISt subsystems