EasyManua.ls Logo

R&S SMA100B - Page 578

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
578User Manual 1178.3834.02 ─ 09
// Configure the Phase modulation settings with combined LF signal sources
SOURce1:PM1:SOURce LF1
SOURce1:PM1:DEViation 1.5
// Select the modulation signal source for the second path
SOURce1:PM2:SOURce LF2
// Combine the signals of both paths with fixed total deviation depth
SOURce1:PM1:DEViation:MODe TOT
SOURce1:PM1:DEViation:SUM 2
// Query the deviation depth of the modulation signal in the second path
SOURce1:PM2:DEViation?
// Response: 0.5
// Combine the signals of both paths with fixed ratio
SOURce1:PM1:DEViation:MODe RAT
SOURce1:PM1:RATio 80
SOURce1:PM2:DEViation?
// Response: 1.2
// Query the deviation depth of modulation signal in the first path
SOURce1:PM1:DEViation?
// Response: 1.5
SOURce1:FM1:STATe 1
SOURce1:FM2:STATe 1
The following commands are available:
[:SOURce<hw>]:PM<ch>:STATe......................................................................................578
[:SOURce<hw>]:PM<ch>:SOURce.................................................................................. 579
[:SOURce<hw>]:PM:MODE............................................................................................ 579
[:SOURce<hw>]:PM:DEViation:MODE............................................................................. 580
[:SOURce<hw>]:PM:DEViation:SUM................................................................................580
[:SOURce<hw>]:PM:RATio..............................................................................................580
[:SOURce<hw>]:PM:SENSitivity?.................................................................................... 580
[:SOURce]:PM<ch>[:DEViation].......................................................................................581
[:SOURce<hw>]:PM<ch>:STATe <State>
Activates phase modulation.
Activation of phase modulation deactivates frequency modulation.
Suffix:
PM<ch>
.
1|2
Sets the modulation signal channel.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Performing phase modulation" on page 577.