613User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:ILS:LOCalizer:COMid:LETTer <Letter>
Sets the length of a Morse code letter space.
Parameters:
<Letter> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.3
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "Letter Space" on page 160
[:SOURce<hw>]:ILS:LOCalizer:COMid:PERiod
<Period>
Sets the period of the COM/ID signal.
Parameters:
<Period> float
Range: 0 to 120
Increment: 1E-3
*RST: 9
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "Period" on page 159
[:SOURce<hw>]:ILS:LOCalizer:COMid:SYMBol <Symbol>
Sets the length of the Morse code symbol space.
Parameters:
<Symbol> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.1
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "Symbol Space" on page 160
[:SOURce<hw>]:ILS:LOCalizer:COMid:TSCHema <TSchema>
Sets the time schema of the Morse code for the COM/ID signal.
Parameters: