Remote control commands
R&S
®
SMA100B
629User Manual 1178.3834.02 ─ 09
Parameters:
<Frequency> float
Range: 0.1 to 20E3
Increment: 0.01
*RST: 1020
Example: See Example "Configure and generate a VOR signal"
on page 596.
Manual operation: See "Frequency" on page 141
[:SOURce<hw>]:VOR:COMid:LETTer
<Letter>
Sets the length of a Morse code letter space.
Parameters:
<Letter> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.3
Default unit: s
Example: See Example "Configure and generate a VOR signal"
on page 596.
Manual operation: See "Letter Space" on page 142
[:SOURce<hw>]:VOR:COMid:PERiod <Period>
Sets the period of the COM/ID signal.
Parameters:
<Period> float
Range: 0 to 120
Increment: 1E-3
*RST: 9
Example: See Example "Configure and generate a VOR signal"
on page 596.
Manual operation: See "Period" on page 141
[:SOURce<hw>]:VOR:COMid:SYMBol <Symbol>
Sets the length of the Morse code symbol space.
Parameters:
<Symbol> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.1
SOURce subsystem