EasyManua.ls Logo

R&S SMA100B - Page 670

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
670User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:LIST:FREQuency:POINts?
Queries the number (points) of frequency entries in the seleced list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Usage: Query only
[:SOURce<hw>]:LIST:INDex <Index>
Sets the list index in LIST:MODE STEP.
After the trigger signal, the instrument processes the frequency and level settings of
the selected index.
Parameters:
<Index> integer
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation: See "Current Index" on page 202
[:SOURce<hw>]:LIST:INDex:STARt <Start>
[:SOURce<hw>]:LIST:INDex:STOP <Stop>
Sets the start and stop index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list.
Parameters:
<Stop> integer
Index range
Only values inside this range are processed in list mode
Range: 0 to list length
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation: See "List Range from/to" on page 205
[:SOURce<hw>]:LIST:MODE <Mode>
Sets the list mode.
SOURce subsystem

Table of Contents

Other manuals for R&S SMA100B

Related product manuals