Remote control commands
R&S
®
SMA100B
671User Manual 1178.3834.02 ─ 09
The instrument processes the list according to the selected mode and trigger source.
See LIST:TRIG:SOUR AUTO, SING or EXT for the description of the trigger
source settings.
Parameters:
<Mode> AUTO | STEP
AUTO
Each trigger event triggers a complete list cycle.
STEP
Each trigger event triggers only one step in the list processing
cycle. The list is processed in ascending order.
*RST: AUTO
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation: See "Mode" on page 203
[:SOURce<hw>]:LIST:POWer
<Power>
Enters the level values in the selected list. The number of level values must corre-
spond to the number of frequency values. Existing data is overwritten.
Parameters:
<Power> <Power#1>{, <Power#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: depends on the installed options
Default unit: dBm
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation: See "Edit List Mode Data" on page 209
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 665.
SOURce subsystem