Remote control commands
R&S
®
SMA100B
672User Manual 1178.3834.02 ─ 09
Usage:  Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:  See Example "Create an RF list and activate the list mode"
on page 665.
Usage:  Event
Manual operation:  See "Execute Single" on page 204
[:SOURce<hw>]:LIST:TRIGger:SOURce 
<Source>
Selects the trigger source for processing lists.
The designation of the parameters correspond to those in sweep mode. SCPI standard
uses other designations for the parameters, which are also accepted by the instrument.
The SCPI designation should be used if compatibility is an important consideration. For
an overview, see the following table:
Rohde & Schwarz parameter SCPI parameter Applies to the list mode parameters:
AUTO IMMediate [:SOURce<hw>]:LIST:MODE AUTO
SINGle BUS
[:SOURce<hw>]:LIST:MODE AUTO or
[:SOURce<hw>]:LIST:MODE STEP
EXTernal EXTernal
[:SOURce<hw>]:LIST:MODE AUTO or
[:SOURce<hw>]:LIST:MODE STEP
Parameters:
<Source> AUTO | IMMediate | SINGle | BUS | EXTernal
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. The selected list is restarted as soon as it is fin-
ished.
SINGle|BUS
The list is triggered by the command [:SOURce<hw>]:LIST:
TRIGger:EXECute. The list is executed once.
EXTernal
The list is triggered externally and executed once.
*RST:  AUTO
Example:  See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation:  See "Mode" on page 203