Remote control commands
R&S
®
SMA100B
702User Manual 1178.3834.02 ─ 09
STEP
Each trigger triggers one sweep step only. The level increases
by the value entered with [:SOURce<hw>]:POWer:STEP[:
INCRement].
*RST: AUTO
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Mode" on page 189
[:SOURce<hw>]:SWEep:POWer:POINts
<Points>
Sets the number of steps within the RF level sweep range.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<Points> integer
Range: 2 to Max
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the level sweep spacing. The sweep spacing for level sweeps is always linear.
Return values:
<Mode> LINear
*RST: LINear
Example:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result: "LIN"
linear spacing
Usage: Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] <Logarithmic>
Sets a logarithmically determined step size for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
SOURce subsystem