Remote control commands
R&S
®
SMA100B
733User Manual 1178.3834.02 ─ 09
Example:
SYSTem:SRData?
// #3118:SOURce1:ROSCillator:SOURce EXT
// :SOURce1:FREQuency:CW 4000000000
// :SOURce1:FREQuency:OFFSet 1000000
// :SOURce1:AM1:STATe 1
// :OUTPut1:STATe 1
Usage: Query only
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete> 1 | ON | 0 | OFF
*RST: 0
Example:
SYST:STAR:COMP?
Response: 1
the startup of the instrument is completed.
Usage: Query only
:SYSTem:DATE <Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
This is a password-protected function. Unlock the protection level 1 to access it.
Parameters:
<Year> integer
<Month> integer
Range: 1 to 12
<Day> integer
Range: 1 to 31
Example:
:SYSTem:DATE?
// 2016,05,01
Manual operation: See "Date" on page 766
:SYSTem:NTP:HOSTname <NTPName>
Sets the address of the NTP server. You can enter the IP address, or the hostname of
the time server, or even set up an own vendor zone. See the Internet for more informa-
tion on NTP.
Parameters:
<NTPName> string
Manual operation: See "NTP Address" on page 767
SYSTem subsystem