12 Injection Port
12.5 Split/Splitless Injection System
137
GC-2010 Plus
12.5.8.2 Parameter list
FLOW RATE
Refer to Fig. 3.5.3, Default:
1
3.0 ml/min
Set the initial flow rate of the flow rate program and the final flow rate for each stage.
Even if no program is set, specify septum purge flow rate in the initial flow rate column.
TIME
Range: 0.00−9999.99 min, Default: 0.00 min
Set the initial flow rate for the flow rate program and the hold time of the final pressure
at each stage.
RATE
Range: END/-400.00−400.00 kPa/min
2
, Default: END
Set the program rate for the flow rate program.
If you set the rate to “0”, “END” appears and the program finishes at the previous ramp.
If you move the cursor to “END” and set any numeric value other than “0”, the pressure
and the time for that ramp can be entered.
1 The purge flow rate range is from 0 to the total flow rate subtracted by the column flow rate and the
split flow rate.
12.5.8.3 PF menu
PF menu Description
Reference
section
Del Line Deletes a line at the current cursor position.
――
Ins Line Inserts a line in line at the current cursor position.
――
On/Off Set On for septum purge flow.
――