13 Detector
13.2 Hydrogen flame ionization detector (FID)
153
GC-2010 Plus
13.2.2.3 PF menu
13.2.2.4 Igniting the FID
Q Ignition procedure
(1) When “DETECTOR” is set to “On” and “AUTO IGNITE” is set to “On” on the [SYSTEM]
key main screen.
PF menu Description
Reference
section
Ignite
Reduces the air flow rate, heats filament, and ignites detector when
pressed when “FLAME” was “Off”.
If “AUTO IGNIT” is set to “On” on the [SYSTEM] key screen, the detector is
automatically ignited after each temperature of detector, the control of
which is “On”, reached the setting value.
13.2.2.4
Det Gas Sets the flow rate of makeup gas, hydrogen and air. 13.2.3
Ign. Set
If the Advanced Pressure Control (APC) is installed, you can set automatic
ignition and automatic re-ignition. Default value is “On”.
―――
Next Det
Toggles to the setup screen of another installed detector.
The [DET] key can also be used to switch screens.
―――
On the [DET] key main screen, set the
detector to “On”.
Press [Start GC] (PF menu).
([SYSTEM] key main screen)
GC control starts. The FID ignites once
it has reached the set temperature.
The baseline should indicate whether
ignition was successful.
The message “Ignition finished.” appears,
and the ignition is completed.