13 Detector
13.2 Hydrogen flame ionization detector (FID)
159
GC-2010 Plus
13.2.3.5 Detector gas program parameters
FLOW RATE
The range and the default valve differs for each gas. See “13.2.3.2 Parameter list”.
Set the default flow rate and the final flow rate for each stage of the detector gas flow
rate program.
TIME
Range: 0.00−9999.99 min, Default: 0.00 min
Set the retention time for the initial flow rate and the final flow rate for each ramp of the
detector gas flow rate program.
RATE
Range: END/-400.0−400.0 ml/min
2
, Default: END
Set the program rate of the detector gas flow rate program. If you set the rate to “0”,
“END” appears and the program ends at the previous ramp. Move the cursor to “END”
and set any numeric value other than “0”, to enter the pressure and the time for that
ramp.
13.2.3.6 PF menu
PF menu Description
Reference
section
Make Up Prog Displays the setup screen of the flow rate program for makeup gas.
――
H2 Prog Displays the setup screen of the flow rate program for hydrogen gas.
――
Air Prog Displays the setup screen of the flow rate program for air.
――
Del Line Deletes a line at the current cursor position.
――
Ins Line Inserts a line in line at the current cursor position.
――