16 Special Functions
16.6 GC Configuration
256
GC-2010 Plus
• The range of the primary pressure is not displayed in the value of “bar”, “psi”. Set up it
with the value that consults the following table and transformed into “kPa”.
• The pressure axis on the screen after pressing [MONIT] key is displayed in the value of
“kPa” regardless of the “pressure unit” sets.
• The setting value, measured value of the primary pressure of the diagnosis result of the
standard diagnosis is displayed in the value of “kPa” regardless of the “pressure unit”
sets.
• The display, printing of each log file is displayed in the value of “kPa” regardless of the
“pressure unit” sets.
BUZZER VOLUME
Selection: Off/Low/Mid/Hig, Default: Low
BUZZER TONE
Selection: Lo/Hi, Default: Hi
Use a different tone for different GCs to be able to discern which GC has a problem.
ATMOSPHERIC COMPENSATION
Selection: On/Off, Default: Off
Select “On” to compensate for the atmospheric pressure.
ZERO TO READY
Selection: On/Off, Default: On
Set “On” for, the gas chromatograph to automatically perform zero adjustment when the
GC is ready and the STATUS light illuminate in green.
POLARITY IN READY
Selection: Open/Close, Default: Open
Set the polarity of GC ready signal.
INJ1/INJ2 PRIMARY PRESS
Selection: 300−500 kPa/500−900 kPa/900−980 kPa, Default: 500−900 kPa
Select the pressure range of the carrier gas cylinder.
Example: If the gas cylinder pressure supplied to the GC-2010 Plus is 700 kPa, select
“500−900 kPa”.
GC START
Selection: SYSTEM Key Screen/ SET Key Screen/ MONIT Key Screen/,
Default: SYSTEM Key Screen
Select the screen, which is displayed automatically after pressing the “Start GC” key
(PF1 key of the [SYSTEM] key main screen).
kPa bar psi
300 - 500 3.00 - 5.00 43.5 - 72.5
500 - 900 5.00 - 9.00 72.5 - 130.5
900 - 980 9.00 - 9.80 130.5 - 142.1