16 Special Functions
16.6 GC Configuration
257
GC-2010 Plus
Gas Saver AOC link
Selection: Yes/No Default: No
This function is displayed on the screen only when the AOC is selected for the line.
This function can be used to perform batch analysis using the AOC. To enable this
function, the Gas Saver function must be set to “On”.
No: When the GC enters the READY state after the completion of analysis, the split
ratio will return from the Gas Saver split ratio to the split ratio specified on the [FLOW]
key screen (at analysis start).
Yes: The Gas Saver split ratio is retained even when the GC enters the READY state
after the completion of analysis.
The split ratio will be changed to the split ratio specified on the [FLOW] key screen when
the signal for starting the AOC is sent via the GC at a start of the next analysis. (This is
available only for batch analysis.)
NOTE • When the GC is in the READY state and the split ratio set by Gas Saver is used after the Gas Saver
AOC link function was activated, the FLOW indicator emits yellow light.
• To return the split ratio to the ratio specified on the [FLOW] key screen (at analysis start) when the
GC is in the READY state and the split ratio set by Gas Saver is used after the Gas Saver AOC link
function was activated, press [Restore] (PF menu) on the [SYSTEM] key screen.
When this function is ON, PF3, which usually serves for [Clean Up], serves for [Restore].
Gas Saver auto on (min)
Range: 0/10 - 120 (min) Default: 0 (min)
Use this function to set a period of time to return the split ratio to the Gas Saver split
ratio after the completion of analysis.
This function is displayed on the screen only when the AOC is selected for the line.
This function is activated when the Gas Saver is set to “On.”
0 (min): When the GC enters the READY state after the completion of analysis,
the split ratio will return to the ratio specified on the [FLOW] key screen
(at analysis start). (Gas Saver split ratio will not be applied.)
10 - 120 (min): When the GC enters the READY state after the completion of analysis,
the split ratio will return from the Gas Saver split ratio to the ratio
specified on the [FLOW] key screen (at analysis start). Then, the Gas
Saver split ratio will be automatically applied after the preset time has
elapsed.
Splitless auto off (min)
Range: 0/10 - 120 (min) Default: 0 (min)
Use this function to set a period of time for the Split Vent to automatically open and
create split state after the completion of analysis.
This function is displayed on the screen only when either the AFC or the Manual Splitter
is attached to the GC.
This function is activated when the split mode of the AFC is set to “SPLITLESS.”
0 (min): When the GC enters the READY state after the completion of analysis,
the Split Vent will close and will remain closed. (Stays ready in the
splitless state.)
10 - 120 (min): When the GC enters the READY state after the completion of analysis,
the Split Vent will close to create the splitless state. When the READY
state has lasted for the preset period of time, the Split Vent will
automatically open and will remain open.