18 Maintenance and Inspection
18.9 Inspection and Maintenance: FID
307
GC-2010 Plus
Fig. 18.9.6   Removing the FID jet
Q Installing the FID jet
Tighten the FID jet securely using a hexnut screwdriver
Fig. 18.9.7   Installing FID jet
Reinstall other parts in reverse order.
Hexnut screwdrive
r
Jet
Fix here

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD