Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 171
Command Reference
E4981A SCPI Commands
:CALibration:CABLe[:LENGth]
Syntax :CALibration:CABLe[:LENGth] <numeric>
:CALibration:CABLe[:LENGth]?
Description Sets the measurement cable length.
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Meas Setup] - CORRECTION - CABLE
<numeric>
Description Desired cable length
Range 0 to 2
Initial value 0
Unit m (meter)
Resolution 1