294 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:TEST:HANDler:STATus:NC
Syntax :TEST:HANDler:STATus:NC {ON|OFF|1|0}
Description This command sets the Handler No Contact/Low C Reject signal for test
purpose and can be used to check/troubleshoot the Test pin signal of Handler
interface. (No Query)
Parameters
Related commands :TEST:HANDler:STATus:ALARm on page 291
:TEST:HANDler:STATus:INDex on page 293
:TEST:HANDler:STATus:OVLD on page 295
:TEST:HANDler:STATus:RDYTrig on page 296
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.
Description
ON Sets No Contact/Low C Reject status to ON.
OFF Sets No Contact/Low C Reject status to OFF.