236 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe]:CORRection:CKIT:STAN3:FORMat
Syntax [:SENSe]:CORRection:CKIT:STAN3:FORMat
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}
[:SENSe]:CORRection:CKIT:STAN3:FORMat?
Description Sets the types of parameters used to define the standard for LOAD correction.
Parameters
For details on Cp and Cs, refer to :CALCulate1:FORMat on page 188. For
details on D, Q, Rp, and Rs, refer to :CALCulate2:FORMat on page 191.
Query response {CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}<newline><^END>
Related commands [:SENSe]:CORRection:CKIT:STAN3[:DATA] on page 235
Equivalent key
sequence
[Meas Setup] - CORRECTION - LOAD- Cp
{CPD|CPQ|CPG|CPRP}/ Cs{CSD|CSQ|CSRS}
Description
CPD
(initial value)
Specifies Cp as the primary parameter, D as the secondary parameter.
CPQ Specifies Cp as the primary parameter, Q as the secondary parameter.
CPG Specifies Cp as the primary parameter, G as the secondary parameter.
CPRP Specifies Cp as the primary parameter, Rp as the secondary parameter.
CSD Specifies Cs as the primary parameter, D as the secondary parameter.
CSQ Specifies Cs as the primary parameter, Q as the secondary parameter.
CSRS Specifies Cs as the primary parameter, Rs as the secondary parameter.