240 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe]:CORRection:DATA
Syntax
[:SENSe]:CORRection:DATA {STAN1|STAN2|STAN3},<numeric 1>,<numeric 2>
[:SENSe]:CORRection:DATA? {STAN1|STAN2|STAN3}
Description Sets the correction data for OPEN/SHORT/LOAD correction.
These correction data are measured for the measurement frequency used in
executing the command (set with the :SOURce:FREQuency[:CW] command).
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe] command), this is set as the
correction data for the selected channel when executing the command
(selected with the [:SENSe]:CORRection:MULTiple:CHANnel command).
To activate the set correction data, use the
[:SENSe]:CORRection:OPEN[:STATe],
[:SENSe]:CORRection:SHORt[:STATe], and
[:SENSe]:CORRection:LOAD[:STATe] commands to turn on the
OPEN/SHORT/LOAD correction function.
The transfer format of data read out with this command conforms to the setup
of the :FORMat[:DATA] command.
Parameters
• When specifying STAN1 as parameter 1:
Parameter 1: {STAN1|STAN2|STAN3}
STAN1 Sets or reads out the data for OPEN correction.
STAN2 Sets or reads out the data for SHORT correction.
STAN3 Sets or reads out the data for LOAD correction.
Parameter 2: <numeric 1> Parameter 3: <numeric 2>
Description
Value of the primary parameter
1
.
Value of the secondary
parameter
*1
.
Range -99.9999E9 to 99.9999E9 -99.9999E9 to 99.9999E9
Initial value 0 0
Unit Depends on the type of parameter. Depends on the type of parameter.
1. Use the [:SENSe]:CORRection:CKIT:STAN1:FORMat command to
specify the type of parameter.