188 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate1:FORMat
Syntax :CALCulate1:FORMat {CP|CS}
:CALCulate1:FORMat?
Description Specifies the primary parameter to be measured.
When the secondary parameter has been set to G or RP and the primary
parameter is set to CS, the secondary parameter is automatically changed to
D. Also, when the secondary parameter has been set to RS and the primary
parameter is set to CP, the secondary parameter is automatically changed to
D.
Parameters
Query response {CP|CS}<newline><^END>
Related commands :CALCulate2:FORMat on page 191
Equivalent key
sequence
[Meas Setup]- FUNC
Description
CP (initial value) Specifies the capacitance value measured using the parallel
equivalent circuit model for the primary parameter.
CS Specifies the capacitance value measured using the series
equivalent circuit model for the primary parameter.