Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 181
Command Reference
E4981A SCPI Commands
:CALCulate1:COMParator:MODE
Syntax :CALCulate1:COMParator:MODE {ABS|DEV|PCNT}
:CALCulate1:COMParator:MODE?
Description Determines how to specify the limit range of the primary parameter for the
comparator function.
Parameters
Query response {ABS|DEV|PCNT}<newline><^END>
Related commands :CALCulate1:COMParator:PRIMary:NOMinal on page 184
:CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT] on page 182
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - Delta Mode
Description
ABS
(initial value)
Specifies the limit border value in an absolute value (absolute mode).
DEV
Specifies the border value relative to the reference value
1
as an absolute
value (absolute tolerance mode).
PCNT Specifies the border value relative to the reference value as a
percentage of the reference value
*1
(percent tolerance mode).
1. Use the :CALCulate1:COMParator:PRIMary:NOMinal command to set the refer-
ence value.