Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 281
Command Reference
E4981A SCPI Commands
:SYSTem:FSHift
Syntax :SYSTem:FSHift <numeric>
:SYSTem:FSHift?
Description When you make a measurement with the measurement frequency set to 1
MHz, specifies the shift of the signal frequency (frequency shift value) actually
applied to the DUT relative to 1 MHz as a percentage of 1 MHz.
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value or the maximum value is set.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Meas Setup] - FREQ SHFT- 0%|1%|-1%|2%|-2%
<numeric>
Description The value of the frequency shift.
Range -2 to 2
Initial (Factory set)
value
0
Unit % (percent)
Resolution 1
A fraction below the resolution is rounded off. MAX or MIN can be used to
specify the parameters.
This command is not available in Option 002.
The value of this command is not changed with *RST and :SYSTem:PRESet
and is only changed with Factory Default reset (available only through
Front Panel).