176 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate1:COMParator:COUNt:DATA?
Syntax :CALCulate1:COMParator:COUNt:DATA?
Description Reads out each count value of BIN1 to BIN9, OUT_OF_BINS, and AUX_BIN of
the BIN counter function.
Reads out all BIN count values regardless of the ON/OFF state of each BIN (set
with the :CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe command).
(Query only)
Parameters
Query response {numeric 1},…,{numeric 11}<newline><^END>
Related commands :CALCulate1:COMParator:COUNt:OVLD? on page 179
:CALCulate1:COMParator:COUNt[:STATe] on page 180
:CALCulate1:COMParator:COUNt:CLEar on page 175
:CALCulate1:COMParator:COUNt:MULTiple:DATA? on page 177
:CALCulate1:COMParator[:STATe] on page 187
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe on page 183
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
{numeric 1} The count value of BIN1.
{numeric 2} The count value of BIN2.
{numeric 3} The count value of BIN3.
{numeric 4} The count value of BIN4.
{numeric 5} The count value of BIN5.
{numeric 6} The count value of BIN6.
{numeric 7} The count value of BIN7.
{numeric 8} The count value of BIN8.
{numeric 9} The count value of BIN9.
{numeric 10} The count value of OUT_OF_BINS.
{numeric 11} The count value of AUX_BIN.