Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 233
Command Reference
E4981A SCPI Commands
[:SENSe]:AVERage[:STATe]
Syntax [:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description Turns ON/OFF the averaging function.
Parameters
Query response {1|0}<newline><^END>
Related commands [:SENSe]:AVERage:COUNt on page 232
Equivalent key
sequence
[Meas Setup] - AVG - ON/OFF
[:SENSe]:CORRection:CKIT:STAN1:FORMat
Syntax [:SENSe]:CORRection:CKIT:STAN1:FORMat {GB|CPG}
[:SENSe]:CORRection:CKIT:STAN1:FORMat?
Description Sets the parameter types of the OPEN correction data.
Parameters
Query response {GB|CPG}<newline><^END>
Related commands [:SENSe]:CORRection:DATA on page 240
Equivalent key
sequence
[Meas Setup] - CORRECTION - OPEN - G-B/Cp-G
Description
ON or 1 (initial value) Turns ON the averaging function.
OFF or 0 Turns OFF the averaging function.
Setting the averaging count with the front panel keys will automatically
turn on the function.
Description
GB (initial value) Specifies G as the primary parameter and B as the secondary
parameter.
CPG Specifies Cp as the primary parameter and G as the secondary
parameter.