Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 239
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO
Syntax [:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO {ON|OFF|1|0}
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO?
Description Sets on/off of auto ranging when the load correction is executed. When this
function is set at off, the range which is selected by
[:SENSe][:FIMPedance]:RANGe[:UPPer] is used for the load correction.
For auto ranging when a measurement is executed, use
[:SENSe][:FIMPedance]:RANGe:AUTO
Parameter
Query response {1|0}<newline><^END>
Related Commands [:SENSe]:CORRection:COLLect[:ACQuire] on page 237
[:SENSe][:FIMPedance]:RANGe[:UPPer] on page 266
:SYSTem:PRESet on page 284
Equivalent key
sequence
[Meas Setup] - CORRECTION - LOAD RNG- AUTO/FIX.
The functionality of this command is same as
[:SENSe]:CORRection:COLLect:LOAD:RANGe:AUTO.
Description
ON or 1 (initial
setting)
Measures the load standard with auto range.
OFF or 0 Measures the load standard measurement range which has been
set just before the load standard measurement is performed (auto
range function is invalid).