246 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe]:CORRection:OFFSet:DATA
Syntax [:SENSe]:CORRection:OFFSet:DATA <numeric 1>,<numeric 2>
[:SENSe]:CORRection:OFFSet:DATA?
Description Sets the correction values for the primary parameter and secondary parameter
used in the offset correction function.
These correction data are measured for the measurement frequency used in
executing the command (set with the :SOURce:FREQuency[:CW] command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response {numeric 1},{numeric 2}<newline><^END>
Related commands [:SENSe]:CORRection:OFFSet[:STATe] on page 247
Equivalent key
sequence
[Meas Setup] - CORRECTION - OFFSET - A|B
<numeric 1> <numeric 2>
Description The offset correction value for the
primary parameter.
The offset correction value for the
secondary parameter.
Range -999.999 to 999.999 -99.9999E9 to 99.9999E9
Initial value 0 0
Unit F (farad) Depends on the setup of the
secondary parameter.
The unit of parameter changes depending on the limit range designation
method.