Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 269
Command Reference
E4981A SCPI Commands
:SOURce:VOLTage:MODE
Syntax :SOURce:VOLTage:MODE {CONTinuous|SYNChronous}
:SOURce:VOLTage:MODE?
Description Determines whether to use the synchronous source function (to output the
measurement signal only during measurement) or not (to always output the
measurement signal). The synchronous source function lets you set the source
delay time with the “:TRIGger[:SEQ1]:DELay” on page 303, to suspend the signal
output during the waiting time after a trigger is generated.
Parameters
Query Response {CONT|SYNC}<newline><^END>
Equivalent key
sequence
[Meas Setup] - SYNC SRC- ON|OFF
:STATus:OPERation:CONDition?
Syntax :STATus:OPERation:CONDition?
Description Reads out the value of the Operation Status Condition register. (Query only)
Query response {numeric}<newline><^END>
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
CONTinuous Always outputs the measurement signal.
SYNChronous Outputs the measurement signal only during measurement.