Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 237
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:COLLect[:ACQuire]
Syntax [:SENSe]:CORRection:COLLect[:ACQuire] {STAN1|STAN2|STAN3}
Description Measures the correction data for OPEN/SHORT/LOAD correction and turns on
the correction function.
For Open and Short correction, the correction data is measured for all
measurement frequencies. For load correction, the correction data is measured
for the measurement frequency used in executing the command (set with the
:SOURce:FREQuency[:CW] command).
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe] command), this is measured as
the correction data for the selected channel when executing the command
(selected with the [:SENSe]:CORRection:MULTiple:CHANnel command).
(No query)
Parameters
Related commands [:SENSe]:CORRection:OPEN[:STATe] on page 248
[:SENSe]:CORRection:SHORt[:STATe] on page 249
[:SENSe]:CORRection:LOAD[:STATe] on page 242
[:SENSe]:CORRection:MULTiple[:STATe] on page 245
[:SENSe]:CORRection:MULTiple:CHANnel on page 243
Equivalent key
sequence
[Meas Setup] - CORRECTION - OPEN - MEAS OPEN
[Meas Setup] - CORRECTION - SHORT - MEAS SHORT
[Meas Setup] - CORRECTION - LOAD - MEAS LOAD
Description
STAN1 Specifies the OPEN correction.
STAN2 Specifies the SHORT correction.
STAN3 Specifies the LOAD correction.