Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 289
Command Reference
E4981A SCPI Commands
:TEST:HANDler:COMP
Syntax :TEST:HANDler:COMP {PHI|PLO|SREJ|OFF}
Description This command sets the Handler Comparator Function value for test purpose
and can be used to check/troubleshoot the Test pin signal of Handler
interface. (No query)
Related commands :TEST:HANDler:BIN on page 288
:TEST:HANDler:KEYLock? on page 289
:TEST:HANDler:MODE on page 290
Equivalent key
sequence
No equivalent key is available on the front panel.
:TEST:HANDler:KEYLock?
Syntax :TEST:HANDler:KEYLock?
Description This command gets the Handler/Key_Lock signal level for test purpose and
can be used to check/troubleshoot the Test pin signal of Handler interface.
(Query only)
Query Response {1|0}<newline><^END>
Related commands :TEST:HANDler:BIN on page 288
:TEST:HANDler:COMP on page 289
:TEST:HANDler:MODE on page 290
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.