276 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CONFigure {AUTO|MANual}
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure?
Description Sets the IP configuration setup method i.e. Auto/Manual.
Parameters
Query response {AUTO|MAN}<newline><^END>
Equivalent key
sequence
[System] - SYSTEM CONFIG - IP CONFIG - AUTO|MANUAL
:SYSTem:COMMunicate:LAN[:SELF]:CONTrol?
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CONTrol?
Description Returns the SOCKET control port number. If the parser is a SOCKET, it will
return a number from 5000 to 5100. Otherwise, 0. (Query Only)
Query response {numeric}<newline><^END>
Equivalent key No equivalent key is available on the front panel.
:SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess?
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess?
Description Returns the current IP address. (Query Only)
Query response {“string”}<newline><^END>
Equivalent key No equivalent key is available on the front panel.
Description
AUTO (Initial [Factory set]
value)
Sets the IP configuration setup to auto mode.
MANual Sets the IP configuration setup to manual mode.
The value of this command is not changed with *RST and :SYSTem:PRESet
and is only changed with Factory Default reset (available only through
Front Panel).