54 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Preparation for Accurate Measurement (Executing Correction)
Using the Multi-correction Function
4-
Using the Multi-correction Function
Turning ON/OFF the multi-correction function
To turn ON/OFF the multi-correction function, use the following command.
[:SENSe]:CORRection:MULTiple[:STATe] on page 245
Selecting a channel
To select a channel for the multi correction function, use the following
command.
[:SENSe]:CORRection:MULTiple:CHANnel on page 243
Measuring correction data
Selecting the definition method of the LOAD correction standard
To select whether to define the LOAD correction standard value (LOAD
correction reference value) for each channel individually or for all channels
commonly, use the following command.
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe] on page 244
Measuring correction data
The method to measure the OPEN/SHORT/LOAD correction data for
multi-correction is the same as that for basic correction data except that you
need to select the proper channel before measurement. For more details, refer
to “Measuring correction data” on page 50.
Reading/writing correction data
The method used to read/write the OPEN/SHORT/LOAD correction data for
multi-correction is the same as that for basic correction data except that you
need to select the proper channel before measurement. For more details, refer
to “Reading/writing correction data (saving/recalling correction conditions)”
on page 51.
Sample program
See Example 9-4 to view an example of multi-correction.