Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 185
Command Reference
E4981A SCPI Commands
:CALCulate1:COMParator:SECondary:LIMit
Syntax :CALCulate1:COMParator:SECondary:LIMit <numeric 1>,<numeric 2>
:CALCulate1:COMParator:SECondary:LIMit?
Description Sets the limit range for the secondary parameter used in the comparator
function.
This command only sets the limit range. To activate the set limit range, use the
:CALCulate1:COMParator:SECondary:STATe command to enable sorting
judgment for the measurement result of the secondary parameter.
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response {numeric 1},{numeric 2}<newline><^END>
Related commands :CALCulate1:COMParator:SECondary:STATe on page 186
:CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT] on page 182
:CALCulate1:COMParator[:STATe] on page 187
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - BIN|2nd - LOW, HIGH
<numeric 1> <numeric 2>
Description The lower limit value of the limit
range.
The upper limit value of the
limit range.
Range -99.9999E9 to 99.9999E9 -99.9999E9 to 99.9999E9
Initial value 0 0
Unit Depends on the type of
parameter.
Depends on the type of
parameter.