Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 191
Command Reference
E4981A SCPI Commands
:CALCulate2:FORMat
Syntax :CALCulate2:FORMat {D|Q|G|RP|RS}
:CALCulate2:FORMat?
Description Specifies the secondary parameter to be measured.
If the primary parameter has been set to CP and the secondary parameter is
set to RS, the primary parameter is automatically changed to CS. Also, if the
primary parameter has been set to CS and the secondary parameter is set to G
or RP, the primary parameter is automatically set to CP.
Parameters
Query response {D|Q|G|RP|RS}<newline><^END>
Related commands :CALCulate1:FORMat on page 188
Equivalent key
sequence
[Meas Setup] - FUNC - Cp{D|Q|G|Rp}|Cs{D|Q|Rs}
Description
D (initial value) Specifies the dissipation factor as the secondary parameter.
Q Specifies the quality factor (inverse value of D) as the secondary
parameter.
G Specifies the equivalent parallel conductance measured using the
parallel equivalent circuit model as the secondary parameter.
RP Specifies the equivalent parallel resistance measured using the
parallel equivalent circuit model as the secondary parameter.
RS Specifies the equivalent series resistance measured using the
series equivalent circuit model as the secondary parameter.