Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Programming Manual
4 Preparation for Accurate Measurement (Executing
Correction)
This chapter describes how to execute the correction function.
Executing OPEN/SHORT/LOAD Correction
Turning ON/OFF correction function
To turn ON/OFF each type of correction, use the corresponding command
below.
Executing the measurement of correction data with the
[:SENSe]:CORRection:COLLect[:ACQuire] command on page 237
automatically measures data and turns ON the correction.
The measured value and the standard value is available to display or input
using the following parameters:
Type of correction Command
OPEN correction
[:SENSe]:CORRection:OPEN[:STATe] on
page 248
SHORT correction
[:SENSe]:CORRection:SHORt[:STATe] on
page 249
LOAD correction
[:SENSe]:CORRection:LOAD[:STATe] on
page 242
Type of correction Parameter
OPEN correction G-B, Cp-G
SHORT correction R-X, Ls-Rs
LOAD correction Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs