208 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:DATA:REFerence2:FILL
Syntax :DATA:REFerence2:FILL
Description Execute a single measurement and set two measured value into each of the
reference values for deviation. This command is same as
:DATA:REFerence1:FILL on page 207. (No Query)
Related commands :DATA:REFerence1:FILL on page 207
Equivalent key
sequence
[Meas Setup] - REF B- MEASURE
:DATA[:DATA]
Syntax :DATA[:DATA] {VMON|IMON|BUF[1-3]|REF1|REF2}
:DATA[:DATA]?
Description Reads out the current monitor value or voltage monitor value of the measured
signal or the data in data buffer 1, data buffer 2 or data buffer 3. Executing this
command rewinds the pointer to the specified data buffer (the location to feed
measurement data) to the start.
The transfer format of data read out with this command conforms to the setup
made with the :FORMat[:DATA] command.
Parameters
Description
IMON Reads out the current monitor.
VMON Reads out the voltage monitor.
BUF1 Reads out data in data buffer 1.
BUF2 Reads out data in data buffer 2.
BUF3 Reads out data in data buffer 3.
Description
REF1 Sets or reads out the reference value for the primary parameter
used in the deviation measurement mode.
REF2 Sets or reads out the reference value for the secondary parameter
used in the deviation measurement mode.