Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 261
Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1
Syntax [:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 <numeric>
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1?
Description Sets/Gets the Threshold1 value for Contact check. Threshold1 is a parameter
which is proportional contact resistance of Hp or Hc. It is recommended to use
the default (initial) value.
Parameters
Query response <numeric><newline><^END>
Related commands [:SENSe][:FIMPedance]:CONTact1:VERify[:STATe] on page 260
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 on page 262
Equivalent key
sequence
[Meas Setup] - CONT CHK- CC1 TH1
The available range for Hp, Hc and Lc is 220 µF to 1 mF at 120 Hz and
22 µF to 100 µF at 1 kHz. The contact resistance for Lp cannot be detected
by contact check feature for this range.
<numeric>
Description Threshold1 value for Contact1.
Range 0 to 1
Initial value 0.1
Resolution 0.01