78 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Sorting Based on Measured Result (Comparator Function)
Reading Out Sorting Judgment Result
7-
Reading Out Sorting Judgment Result
You can read out the sorting judgment result (readout value) along with the
measured result by the comparator function according to the correspondences
in Figure 7-1. To do read out sorting judgment results, use the following
commands:
• *TRG on page 163
• :FETCh? on page 220
• :READ? on page 230
The comparator sorting result is read out as an integer between 0 and 11, as
shown in the table below.
To display the comparator sorting result use the following command to select
the <BIN No. DISPLAY> page.
• :DISPlay:PAGE on page 213
Table 7-2 Relationship between comparator sorting result and readout value
Readout value Comparator sorting result
0OUT_OF_BINS
1BIN1
2BIN2
3BIN3
4BIN4
5BIN5
6BIN6
7BIN7
8BIN8
9BIN9
10 AUX_BIN
11 Cannot be sorted (OVLD or NC is detected)