Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 53
Preparation for Accurate Measurement (Executing Correction)
Executing Offset Correction
Executing Offset Correction
Turning ON/OFF the correction function
When you turn ON the offset correction, assuming that the measured value
before correction is and the offset correction data is , the measured
value is compensated as .
To turn ON/OFF the offset correction, use the following command.
[:SENSe]:CORRection:OFFSet[:STATe] on page 247
You cannot turn ON/OFF the primary parameter and secondary parameter
separately. However, if you set the correction value to 0, the state is actually
the same as OFF even if the offset correction is ON. Therefore, in practice you
can realize separate ON/OFF states by setting the correction value for either
parameter to 0.
Setting up the correction data
To set up the offset correction data, use the following command.
[:SENSe]:CORRection:OFFSet:DATA on page 246
The entered value is set as the offset correction data for the measurement
frequency at the time of the entry.
Meas
Offset
Meas Offset