282 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:SYSTem:HANDler:TRIGger:VOLTage
Syntax :SYSTem:HANDler:TRIGger:VOLTage <numeric>
:SYSTem:HANDler:TRIGger:VOLTage?
Description Sets the handler trigger input voltage.
Parameters
Query response {numeric}<newline><^END>
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description The value of the handler trigger input voltage
Range 5 to 24
Initial (Factory set)
value
24
Unit V (voltage)
Resolution 100m
The value of this command is not changed with *RST and :SYSTem:PRESet
and is only changed with Factory Default reset (available only through
Front Panel).
The product may get damaged if incorrect voltage is set to Handler
Trigger. Take caution while changing the Handler Trigger voltage.