Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 187
Command Reference
E4981A SCPI Commands
:CALCulate1:COMParator[:STATe]
Syntax :CALCulate1:COMParator[:STATe] {ON|OFF|1|0}
:CALCulate1:COMParator[:STATe]?
Description Turns ON/OFF the comparator function.
This setup is interlocked with the ON/OFF state of the signal output of the
handler interface.
Parameters
Query response {1|0}<newline><^END>
Related commands :CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - COMP
Changing the measurement parameter will automatically turn off this
function. To avoid this, you have to write a program so that this command
is executed after the execution of the measurement parameter setup
command (:CALCulate1:FORMat and :CALCulate2:FORMat).
Description
ON or 1 Turns ON the comparator function.
OFF or 0 (initial value) Turns OFF the comparator function.