Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 291
Command Reference
E4981A SCPI Commands
:TEST:HANDler:STATus:ALARm
Syntax :TEST:HANDler:STATus:ALARm {ON|OFF|1|0}
Description This command sets the Handler Alarm signal for test purpose and can be used
to check/troubleshoot the Test pin signal of Handler interface. (No Query)
Parameters
Related commands :TEST:HANDler:STATus:INDex on page 293
:TEST:HANDler:STATus:NC on page 294
:TEST:HANDler:STATus:OVLD on page 295
:TEST:HANDler:STATus:RDYTrig on page 296
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.
Description
ON Sets Handler alarm signal to ON.
OFF Sets Handler alarm signal to OFF.