Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 227
Command Reference
E4981A SCPI Commands
:INITiate:CONTinuous
Syntax :INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description This command changes the trigger state from “idle” state to the “wait for
trigger” automatically or stay in “idle” state. After executing :SYSTem:PRESet,
this value changes to ON, after *RST this value changes to OFF .
For details on the trigger system, refer to “Trigger system” on page 57.
Parameters
This setup is initialized to ON when using the :SYSTem:PRESet command and
to OFF when using the *RST command.
Query response {1|0}<newline><^END>
Related commands *RST on page 160
:SYSTem:PRESet on page 284
Equivalent key
sequence
No equivalent key is available on the front panel.
This command is turned ON when the trigger source is changed from the
front panel of the E4981A.
Description
ON or 1 Enables successive startup.