Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 297
Command Reference
E4981A SCPI Commands
:TEST:HANDler:TRIGger?
Syntax :TEST:HANDler:TRIGger?
Description This command gets the Handler Trigger signal for test purpose and can be
used to check/troubleshoot the Test pin signal of Handler interface. (Query
Only)
Query response {1|0}<newline><^END>
Related commands :TEST:REAR:TRIGger? on page 297
Equivalent key
sequence
No equivalent key is available on the front panel.
:TEST:REAR:TRIGger?
Syntax :TEST:REAR:TRIGger?
Description This command gets the BNC Trigger signal for test purpose and can be used to
check/troubleshoot the Test pin signal of Handler interface. (Query Only)
Query response {1|0}<newline><^END>
Related commands :TEST:HANDler:TRIGger? on page 297
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.