Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 183
Command Reference
E4981A SCPI Commands
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe
Syntax :CALCulate1:COMParator:PRIMary:BIN{1|2|3|4|5|6|7|8|9}:STATe {ON|OFF|1|0}
:CALCulate1:COMParator:PRIMary:BIN{1|2|3|4|5|6|7|8|9}:STATe?
Description Turns ON/OFF BIN1 to BIN9 of the comparator function.
Only BINs that you turn on using this command are used for the sorting
judgment of the comparator function.
Parameters
Query response {1|0}<newline><^END>
Related commands :CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT] on page 182
:CALCulate1:COMParator:SECondary:STATe on page 186
:CALCulate1:COMParator[:STATe] on page 187
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - BIN{1|2|3|4|5|6|7|8|9}- ON|OFF
Description
ON or 1
(initial value of BIN1)
Turns ON BIN.
OFF or 0
(initial value of BIN2 to BIN9)
Turns OFF BIN.