306 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:TRIGger:SEQ2:DELay
Syntax :TRIGger:SEQ2:DELay <numeric>[mS|S]
:TRIGger:SEQ2:DELay?
Description Specifies the waiting time between when a trigger is inputted and when the
measurement starts (trigger delay time).
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Meas Setup] - TRIG DLY
<numeric>
Description Trigger delay time
Range 0 to 1
Initial value 0
Unit s (second)
Resolution 100u
A fraction below the resolution is rounded off. MAX or MIN can be used to
specify the parameters.