228 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:INITiate[:IMMediate]
Syntax :INITiate[:IMMediate]
Description Changes the trigger state from “idle” state to “wait for trigger” state for one
trigger sequence. (No Query)
For details on the trigger system, refer to “Trigger system” on page 57. (No
query)
Related commands :INITiate:CONTinuous on page 227
Equivalent key
sequence
No equivalent key is available on the front panel.
:MMEMory:DELete[:REGister]
Syntax :MMEMory:DELete[:REGister] <numeric>
Description Deletes the state from the memories. Numbers 0 to 9 are located in the
internal memory, while No. 10 to 19 are situated in the external USB memory.
(No Query)
Parameter
Equivalent key [Save/Recall] - No {1-9} - DELETE
<Numeric>
Range 0 to 19
Resolution 1

Table of Contents