242 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
Equivalent key
sequence
[Meas Setup] - CORRECTION - OPEN - A|B
[Meas Setup] - CORRECTION - SHORT- A|B
[Meas Setup] - CORRECTION - LOAD- A|B
[:SENSe]:CORRection:LOAD[:STATe]
Syntax [:SENSe]:CORRection:LOAD[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:LOAD[:STATe]?
Description Turns ON/OFF the LOAD correction function.
With the LOAD correction set to ON, if you change the cable length (set with
the :CALibration:CABLe[:LENGth] command) or frequency shift (set with
the :SYSTem:FSHift command), the LOAD correction is automatically
changed to OFF.
Parameters
Query response {1|0}<newline><^END>
Related commands :CALibration:CABLe[:LENGth] on page 171
:SYSTem:FSHift on page 281
[:SENSe]:CORRection:COLLect[:ACQuire] on page 237
Equivalent key
sequence
[Meas Setup] - CORRECTION - LOAD - ON/OFF
Description
ON or 1 Turns ON the LOAD correction.
OFF or 0 (initial value) Turns OFF the LOAD correction.