300 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:TEST:SCANner:INDex
Syntax :TEST:SCANner:INDex {ON|OFF|1|0}
Description This command sets the Scanner Index signal for test purpose and can be used
to check/troubleshoot the Test pin signal of Scanner interface. (No Query)
Parameters
Related commands :TEST:SCANner:CH? on page 298
:TEST:SCANner:EOM on page 299
:TEST:SCANner:MODE on page 301
:TEST:SCANner:TRIGger? on page 302
:TEST:SCANner:VALid? on page 302
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:SCANner:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:SCAN:xxxx commands. After completing
the test, :TEST:SCANner:MODE should set to OFF so that scanner interface
pin signal is generated according to the actual measurement result.
Description
ON Sets Scanner Index signal status to ON.
OFF Sets Scanner Index status to OFF.