Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 205
Command Reference
E4981A SCPI Commands
:DATA:POINts[:DATA]
Syntax :DATA:POINts[:DATA] {BUF1|BUF2|BUF3},<numeric>
:DATA:POINts[:DATA]? {BUF1|BUF2|BUF3}
Description Specifies the size of data buffer 1, data buffer 2, or data buffer 3 in number of
measurements. Executing this command returns the pointer to the specified
data buffer (the location to feed measurement data) back to the start.
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response {numeric}<newline><^END>
Related commands :DATA:FEED:BUF1 on page 195
:DATA:FEED:CONTrol[:STATe] on page 200
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
BUF1 Specifies data buffer 1 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 1.
BUF2 Specifies data buffer 2 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 2.
BUF3 Specifies data buffer 3 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 3.
<numeric>
Description Number of measurements for the data buffer specified with
{BUF1|BUF2|BUF3}
Range Data buffer 1: 1 to 200 Data
buffer 2: 1 to 200 Data buffer
3: 1 to 1000
Initial value Data buffer 1: 200 Data
buffer 2: 200 Data buffer
3: 1000
Resolution 1